4Z78
Weak TCR binding to an unstable insulin epitope drives type 1 diabetes
Experimental procedure
Experimental method | SINGLE WAVELENGTH |
Source type | SYNCHROTRON |
Source details | DIAMOND BEAMLINE I04-1 |
Synchrotron site | Diamond |
Beamline | I04-1 |
Temperature [K] | 100 |
Detector technology | CCD |
Collection date | 2012-02-09 |
Detector | ADSC QUANTUM 315r |
Wavelength(s) | 0.9173 |
Spacegroup name | P 21 21 21 |
Unit cell lengths | 46.170, 151.570, 182.230 |
Unit cell angles | 90.00, 90.00, 90.00 |
Refinement procedure
Resolution | 78.090 - 2.304 |
R-factor | 0.1904 |
Rwork | 0.188 |
R-free | 0.23270 |
Structure solution method | MOLECULAR REPLACEMENT |
RMSD bond length | 0.015 |
RMSD bond angle | 1.744 |
Data scaling software | SCALA (3.3.20) |
Phasing software | PHASER (2.5.6) |
Refinement software | REFMAC (5.8.0073) |
Data quality characteristics
Overall | Inner shell | Outer shell | |
Low resolution limit [Å] | 182.230 | 182.230 | 2.360 |
High resolution limit [Å] | 2.304 | 10.310 | 2.300 |
Rmerge | 0.041 | 0.929 | |
Rmeas | 0.148 | ||
Rpim | 0.055 | 0.019 | 0.399 |
Total number of observations | 418482 | 4699 | 30238 |
Number of reflections | 57725 | ||
<I/σ(I)> | 10.4 | 21.3 | 3 |
Completeness [%] | 100.0 | 99.8 | 100 |
Redundancy | 7.2 | 6 | 7.3 |
Crystallization Conditions
crystal ID | method | pH | temperature | details |
1 | VAPOR DIFFUSION | 291 | 4% PEG 4000, 0.1 M sodium acetate pH 4.6 |