Experimental procedure
Experimental method | MAD |
Source type | SYNCHROTRON |
Source details | SOLEIL BEAMLINE PROXIMA 1 |
Synchrotron site | SOLEIL |
Beamline | PROXIMA 1 |
Temperature [K] | 100 |
Detector technology | PIXEL |
Collection date | 2017-07-11 |
Detector | DECTRIS EIGER X 16M |
Wavelength(s) | 0.9794 |
Spacegroup name | P 1 21 1 |
Unit cell lengths | 63.710, 76.850, 178.650 |
Unit cell angles | 90.00, 97.30, 90.00 |
Refinement procedure
Resolution | 48.810 - 2.640 |
R-factor | 0.2594 |
Rwork | 0.259 |
R-free | 0.26410 |
Structure solution method | SAD |
RMSD bond length | 0.007 |
RMSD bond angle | 0.870 |
Data reduction software | XDS |
Data scaling software | Aimless |
Phasing software | SHELXCD |
Refinement software | BUSTER (2.10.4 (3-FEB-2022)) |
Data quality characteristics
Overall | Outer shell | |
Low resolution limit [Å] | 48.810 | 2.710 |
High resolution limit [Å] | 2.640 | 2.640 |
Number of reflections | 45042 | 2252 |
<I/σ(I)> | 16 | |
Completeness [%] | 89.0 | |
Redundancy | 13.6 | |
CC(1/2) | 0.998 | 0.429 |
Crystallization Conditions
crystal ID | method | pH | temperature | details |
1 | VAPOR DIFFUSION, SITTING DROP | 293 | 0.2M Lithium chloride 24% PEG 3350 0.1M MES pH 5.5 |