7B0H

TgoT_6G12 Ternary complex

Experimental procedure
Experimental methodSINGLE WAVELENGTH
Source typeSYNCHROTRON
Source detailsSOLEIL BEAMLINE PROXIMA 1
Synchrotron siteSOLEIL
BeamlinePROXIMA 1
Temperature [K]293
Detector technologyPIXEL
Collection date2019-07-15
DetectorDECTRIS EIGER X 16M
Wavelength(s)0.987
Spacegroup nameP 21 21 21
Unit cell lengths111.823, 112.331, 186.518
Unit cell angles90.00, 90.00, 90.00
Refinement procedure
Resolution48.920 - 3.150
R-factor0.2327
Rwork0.232
R-free0.24720
Structure solution methodMOLECULAR REPLACEMENT
Starting model (for MR)1tgo
RMSD bond length0.005
RMSD bond angle0.780
Data reduction softwareXDS
Data scaling softwareXSCALE
Phasing softwarePHENIX
Refinement softwareBUSTER (2.10.3 (20-MAY-2020))
Data quality characteristics
 OverallOuter shell
Low resolution limit [Å]48.9203.263
High resolution limit [Å]3.1503.150
Number of reflections333171758
<I/σ(I)>17.89
Completeness [%]80.5
Redundancy2.7
CC(1/2)0.9970.530
Crystallization Conditions
crystal IDmethodpHtemperaturedetails
1VAPOR DIFFUSION, HANGING DROP2935% PEG 20K and 0.1M MES-NaOH pH 6.0

238895