Experimental procedure
| Experimental method | SINGLE WAVELENGTH |
| Source type | SYNCHROTRON |
| Source details | DIAMOND BEAMLINE I23 |
| Synchrotron site | Diamond |
| Beamline | I23 |
| Temperature [K] | 80 |
| Detector technology | PIXEL |
| Collection date | 2019-01-24 |
| Detector | DECTRIS PILATUS 12M |
| Wavelength(s) | 2.755 |
| Spacegroup name | P 3 2 1 |
| Unit cell lengths | 149.421, 149.421, 45.810 |
| Unit cell angles | 90.00, 90.00, 120.00 |
Refinement procedure
| Resolution | 74.823 - 2.300 |
| Rwork | 0.183 |
| R-free | 0.23590 |
| Structure solution method | SAD |
| RMSD bond length | 0.009 |
| RMSD bond angle | 1.645 |
| Data reduction software | XDS |
| Data scaling software | Aimless |
| Phasing software | CRANK2 |
| Refinement software | REFMAC (5.8.0258) |
Data quality characteristics
| Overall | Outer shell | |
| Low resolution limit [Å] | 129.400 | 2.380 |
| High resolution limit [Å] | 2.300 | 2.300 |
| Rmerge | 0.151 | |
| Rpim | 0.023 | 0.154 |
| Number of reflections | 24537 | 2058 |
| <I/σ(I)> | 19 | 4.2 |
| Completeness [%] | 93.6 | 85 |
| Redundancy | 40.9 | 34.2 |
| CC(1/2) | 0.998 | 0.850 |
Crystallization Conditions
| crystal ID | method | pH | temperature | details |
| 1 | VAPOR DIFFUSION, SITTING DROP | 7.5 | 293 | PEG 4000, HEPES |