Experimental procedure
Experimental method | SINGLE WAVELENGTH |
Source type | SYNCHROTRON |
Source details | APS BEAMLINE 19-BM |
Synchrotron site | APS |
Beamline | 19-BM |
Temperature [K] | 100 |
Detector technology | AREA DETECTOR |
Collection date | 2004-06-06 |
Detector | SBC-3 |
Wavelength(s) | 0.97895 |
Spacegroup name | P 43 21 2 |
Unit cell lengths | 56.943, 56.943, 89.983 |
Unit cell angles | 90.00, 90.00, 90.00 |
Refinement procedure
Resolution | 30.004 - 1.369 |
R-factor | 0.1488 |
Rwork | 0.148 |
R-free | 0.16900 |
Structure solution method | MOLECULAR REPLACEMENT |
Starting model (for MR) | 3KWT |
Data reduction software | HKL-2000 |
Data scaling software | HKL-2000 |
Phasing software | AMoRE |
Refinement software | PHENIX (1.14_3260: ???) |
Data quality characteristics
Overall | Inner shell | Outer shell | |
Low resolution limit [Å] | 50.000 | 30.004 | 1.390 |
High resolution limit [Å] | 1.369 | 3.720 | 1.370 |
Rmerge | 0.055 | 0.028 | 0.464 |
Number of reflections | 31875 | 1784 | 1483 |
<I/σ(I)> | 15.3 | ||
Completeness [%] | 99.8 | 99.2 | 96.4 |
Redundancy | 11.6 | 11.6 | 5.9 |
Crystallization Conditions
crystal ID | method | pH | temperature | details |
1 | VAPOR DIFFUSION, HANGING DROP | 6.8 | 298 | 30% PEG-MME 2000, 0.1 M bis-tris propane pH 6.8, 0.1 M NaCl, 0.1 MM CaCl2, 0.5 mM TCEP |