4MTQ

Ni-bound GloA2

Experimental procedure
Experimental methodSINGLE WAVELENGTH
Source typeSYNCHROTRON
Source detailsAUSTRALIAN SYNCHROTRON BEAMLINE MX2
Synchrotron siteAustralian Synchrotron
BeamlineMX2
Temperature [K]100
Detector technologyCCD
Collection date2010-10-26
DetectorADSC QUANTUM 315r
Wavelength(s)0.95369
Spacegroup nameP 21 21 21
Unit cell lengths41.145, 74.983, 80.362
Unit cell angles90.00, 90.00, 90.00
Refinement procedure
Resolution19.410 - 2.170
R-factor0.1853
Rwork0.183
R-free0.23290
Structure solution methodMOLECULAR REPLACEMENT
Starting model (for MR)4mtr
RMSD bond length0.010
RMSD bond angle1.020
Data reduction softwareXDS
Data scaling softwareSCALA
Phasing softwareMOLREP
Refinement softwareBUSTER (2.10.0)
Data quality characteristics
 OverallOuter shell
Low resolution limit [Å]19.8402.280
High resolution limit [Å]2.1702.170
Rmerge0.0760.688
Number of reflections13652
<I/σ(I)>16.12.2
Completeness [%]98.898.8
Redundancy6.64.4
Crystallization Conditions
crystal IDmethodpHtemperaturedetails
1VAPOR DIFFUSION6.32930.1 M succinic acid, 21% PEG3350, pH 6.3, VAPOR DIFFUSION, temperature 293K

224201

PDB entries from 2024-08-28