Loading
PDBj
MenuPDBj@FacebookPDBj@TwitterPDBj@YouTubewwPDB FoundationwwPDB
RCSB PDBPDBeBMRBAdv. SearchSearch help

2OLQ

How Does an Enzyme Recognize CO2?

Experimental procedure
Experimental methodSINGLE WAVELENGTH
Source typeSYNCHROTRON
Source detailsAPS BEAMLINE 14-BM-D
Synchrotron siteAPS
Beamline14-BM-D
Temperature [K]100
Detector technologyCCD
Collection date2003-11-17
DetectorADSC QUANTUM 4
Spacegroup nameC 1 2 1
Unit cell lengths125.584, 91.369, 46.577
Unit cell angles90.00, 96.22, 90.00
Refinement procedure
Resolution50.000 - 1.940
R-factor0.204
Rwork0.205
R-free0.23500
Structure solution methodMOLECULAR REPLACEMENT
Starting model (for MR)1ayl
Data reduction softwareDENZO
Data scaling softwareSCALEPACK
Phasing softwareAMoRE
Refinement softwareCNS
Data quality characteristics
 OverallOuter shell
Low resolution limit [Å]50.0002.040
High resolution limit [Å]1.7201.940
Rmerge0.0660.516
Number of reflections52927
<I/σ(I)>7.41.5
Completeness [%]95.676.1
Redundancy3.53.1
Crystallization Conditions
crystal IDmethodpHtemperaturedetails
1VAPOR DIFFUSION4.529830% P4000, 0.1 M sodium acetate pH 4.5, 0.2 M ammonium acetate, VAPOR DIFFUSION, temperature 298K

226707

PDB entries from 2024-10-30

PDB statisticsPDBj update infoContact PDBjnumon