Loading
PDBj
MenuPDBj@FacebookPDBj@X(formerly Twitter)PDBj@BlueSkyPDBj@YouTubewwPDB FoundationwwPDBDonate
RCSB PDBPDBeBMRBAdv. SearchSearch help

9EQV

41fs pulse duration 10uJ pulse energy thaumatin

Experimental procedure
Experimental methodSINGLE WAVELENGTH
Source typeFREE ELECTRON LASER
Source detailsSwissFEL ARAMIS BEAMLINE ESC
Synchrotron siteSwissFEL ARAMIS
BeamlineESC
Temperature [K]293
Detector technologyPIXEL
Collection date2023-09-05
DetectorPSI JUNGFRAU 8M
Wavelength(s)1.0306
Spacegroup nameP 41 21 2
Unit cell lengths58.500, 58.500, 151.250
Unit cell angles90.00, 90.00, 90.00
Refinement procedure
Resolution32.000 - 1.480
Rwork0.167
R-free0.20050
Structure solution methodMOLECULAR REPLACEMENT
RMSD bond length0.010
RMSD bond angle1.722
Data reduction softwareCrystFEL (10.2)
Data scaling softwareCrystFEL (10.2)
Phasing softwarePHASER
Refinement softwareREFMAC (5.8.0425)
Data quality characteristics
 OverallOuter shell
Low resolution limit [Å]32.0001.510
High resolution limit [Å]1.4801.480
Number of reflections457532221
<I/σ(I)>4.260.71
Completeness [%]100.0100
Redundancy122.2
CC(1/2)0.9500.357
Crystallization Conditions
crystal IDmethodpHtemperaturedetails
1BATCH MODE293100 mg of thaumatin (Sigma-T7638) was dissolved in 1 ml of MiniQ water, reaching a concentration of 100 mg/ml with solutions vortexed to ensure complete dissolution. Subsequently, 200 uL of the thaumatin solution was mixed with 200 ul of 1.6 M (Na, K, tartrate) crystallization solution