6K2C
Extended Hect domain of UBE3C E3 Ligase
Experimental procedure
Experimental method | SINGLE WAVELENGTH |
Source type | SYNCHROTRON |
Source details | APS BEAMLINE 24-ID-E |
Synchrotron site | APS |
Beamline | 24-ID-E |
Temperature [K] | 296 |
Detector technology | PIXEL |
Collection date | 2018-09-20 |
Detector | DECTRIS PILATUS3 S 6M |
Wavelength(s) | 0.9791 |
Spacegroup name | P 43 21 2 |
Unit cell lengths | 75.610, 75.610, 220.190 |
Unit cell angles | 90.00, 90.00, 90.00 |
Refinement procedure
Resolution | 48.094 - 2.703 |
R-factor | 0.2392 |
Rwork | 0.237 |
R-free | 0.27730 |
Structure solution method | MOLECULAR REPLACEMENT |
Starting model (for MR) | 2oni |
RMSD bond length | 0.008 |
RMSD bond angle | 0.931 |
Data reduction software | HKL-2000 |
Data scaling software | HKL-2000 |
Phasing software | PHASER |
Refinement software | PHENIX ((1.13_2998: ???)) |
Data quality characteristics
Overall | Outer shell | |
Low resolution limit [Å] | 48.093 | 2.796 |
High resolution limit [Å] | 2.700 | 2.700 |
Rmeas | 0.169 | |
Rpim | 0.036 | 0.106 |
Number of reflections | 18388 | 1780 |
<I/σ(I)> | 32.45 | |
Completeness [%] | 99.9 | |
Redundancy | 20.9 | |
CC(1/2) | 0.992 | 0.981 |
Crystallization Conditions
crystal ID | method | pH | temperature | details |
1 | VAPOR DIFFUSION | 296 | 100mM HEPES 7.5, 20% PEG 10000 |