Loading
PDBj
MenuPDBj@FacebookPDBj@TwitterPDBj@YouTubewwPDB FoundationwwPDB
RCSB PDBPDBeBMRBAdv. SearchSearch help

6GXF

The hit-and-return system enables efficient time-resolved serial synchrotron crystallography: RADDAM1

Experimental procedure
Experimental methodSINGLE WAVELENGTH
Source typeSYNCHROTRON
Source detailsPETRA III, DESY BEAMLINE P11
Synchrotron sitePETRA III, DESY
BeamlineP11
Temperature [K]293
Detector technologyPIXEL
Collection date2017-05-01
DetectorDECTRIS PILATUS 6M
Wavelength(s)1.00890
Spacegroup nameP 1 21 1
Unit cell lengths41.320, 78.540, 83.560
Unit cell angles90.00, 102.92, 90.00
Refinement procedure
Resolution40.722 - 1.800
R-factor0.2243
Rwork0.222
R-free0.26900
Structure solution methodMOLECULAR REPLACEMENT
Starting model (for MR)3r3u
RMSD bond length0.006
RMSD bond angle0.820
Data reduction softwareXDS (nXDS)
Data scaling softwareXSCALE (nXSCALE)
Phasing softwarePHASER
Refinement softwarePHENIX ((1.12_2829: ???))
Data quality characteristics
 OverallOuter shell
Low resolution limit [Å]81.4201.830
High resolution limit [Å]1.8001.800
Number of reflections48178
<I/σ(I)>2.9
Completeness [%]99.8
Redundancy9.2
Crystallization Conditions
crystal IDmethodpHtemperaturedetails
1BATCH MODE293PEG 3350 18 - 20 % CaCl 200 mM TrisHCL 100 mM pH 8.5

219869

PDB entries from 2024-05-15

PDB statisticsPDBj update infoContact PDBjnumon