Loading
PDBj
MenuPDBj@FacebookPDBj@TwitterPDBj@YouTubewwPDB FoundationwwPDB
RCSB PDBPDBeBMRBAdv. SearchSearch help

4H9I

Radiation damage study of lysozyme - 1.05 MGy

Experimental procedure
Experimental methodSINGLE WAVELENGTH
Source typeSYNCHROTRON
Source detailsSSRL BEAMLINE BL9-2
Synchrotron siteSSRL
BeamlineBL9-2
Temperature [K]100
Detector technologyCCD
Collection date2011-02-12
DetectorMARMOSAIC 325 mm CCD
Wavelength(s)1.0332
Spacegroup nameP 43 21 2
Unit cell lengths78.774, 78.774, 36.870
Unit cell angles90.00, 90.00, 90.00
Refinement procedure
Resolution26.917 - 1.200
R-factor0.1918
Rwork0.191
R-free0.20300
Structure solution methodMOLECULAR REPLACEMENT
Starting model (for MR)6lyz
RMSD bond length0.007
RMSD bond angle1.136
Data reduction softwareDENZO
Data scaling softwareSCALEPACK
Phasing softwareMOLREP
Refinement softwarePHENIX (1.7.2_862)
Data quality characteristics
 OverallInner shellOuter shell
Low resolution limit [Å]50.00050.0001.220
High resolution limit [Å]1.2003.2601.200
Rmerge0.0320.0180.316
Number of reflections36619
<I/σ(I)>15.2
Completeness [%]99.393.1100
Redundancy4.54.44.3
Crystallization Conditions
crystal IDmethodpHtemperaturedetails
1VAPOR DIFFUSION, HANGING DROP4.82987.5% sodium chloride, 100 mM sodium acetate pH 4.8, 25% ethylene glycol, vapor diffusion, hanging drop, temperature 298K