+検索条件
-Structure paper
タイトル | Multishot tomography for high-resolution in situ subtomogram averaging. |
---|---|
ジャーナル・号・ページ | J Struct Biol, Vol. 215, Issue 1, Page 107911, Year 2023 |
掲載日 | 2022年11月4日 |
著者 | S Khavnekar / W Wan / P Majumder / W Wietrzynski / P S Erdmann / J M Plitzko / |
PubMed 要旨 | Cryo-electron tomography (cryo-ET) and subtomogram averaging (STA) can resolve protein complexes at near atomic resolution, and when combined with focused ion beam (FIB) milling, macromolecules can ...Cryo-electron tomography (cryo-ET) and subtomogram averaging (STA) can resolve protein complexes at near atomic resolution, and when combined with focused ion beam (FIB) milling, macromolecules can be observed within their native context. Unlike single particle acquisition (SPA), cryo-ET can be slow, which may reduce overall project throughput. We here propose a fast, multi-position tomographic acquisition scheme based on beam-tilt corrected beam-shift imaging along the tilt axis, which yields sub-nanometer in situ STA averages. |
リンク | J Struct Biol / PubMed:36343843 |
手法 | EM (サブトモグラム平均) |
解像度 | 4.7 - 8.8 Å |
構造データ | EMDB-16139: E.coli 70S ribosome subtomogram average from multishot acquisition on cryo-FIB lamellae EMDB-16162: E.coli 70S ribosome subtomogram average from singleshot acquisition on mixed Ribosome-Proteasome sample EMDB-16165: E.coli 70S ribosome subtomogram average from multishot acquisition on mixed Ribosome-Proteasome sample EMDB-16180: 20S Proteasome subtomogram average from Singleshot acquisition on mixed Riobsome-Proteasome sample EMDB-16181: 20S Proteasome subtomogram average from multishot tomography acquisition on mixed Ribosome-Proteasome sample |
由来 |
|